MR Finger (Not Thumb) WO and W/WO Protocol

Revised - 2/22/19
3T Scanner Only

  • Flex/ Head Coil
  • Use rest slab proximal to wrist to reduce flow and pulsatile motion
Plane Weighting Mode Slice Gap FAT SAT FOV Scan Range
AXIAL T1 TSE 3mm 0.5mm None 12cm Cover from tip of finger soft tissue to distal metacarpal.Include finger of interest and adjacent two fingers.
AXIAL T2 FS Mid TE (40-50) TSE 3mm 0.5mm SPAIR 12cm Coverage same as Ax T1.Include finger of interest and adjacent two fingers.
COR T1 TSE 3mm 0.5mm None 12cm Angle parallel to metacarpal head.Include finger of interest and adjacent two fingers.
COR T2 FS Mid TE (40-50)(replace with STIR if for infection) TSE 3mm 0.5mm SPAIR(or STIR) 12cm Coverage same as Cor T1.Include finger of interest and adjacent two fingers
SAG T2 FS Mid TE (40-50)(replace with STIR if for infection) TSE 3mm 0.5mm SPAIR(or STIR) 12cm Perpendicular to coronal plane.Include finger of interest and adjacent two fingers
SAG PRE(if giving gad for infection/osteo) T1 TSE 3mm 0.5mm None 12cm Same as SAG T2 FS
CONTRAST INJECTION
AXIAL T1 Fat Sat TSE 3mm 0.5mm SPAIR 12cm Same as pre
COR T1 Fat Sat TSE 3mm 0.5mm SPAIR 12cm Same as pre
SAG T1 Fat Sat TSE 3mm 0.5mm SPAIR 12cm Same as pre
MR Finger Not Thumb Image