MR Hand WO or W/WO Protocol
Revised - 2/22/19
Charge as Hand WO or WWO
- 3T Scanner Only
- Flex/ Head Coil
- Use REST slab proximal to wrist to reduce flow and pulsatile motion
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Scan Range |
---|---|---|---|---|---|---|---|
AXIAL | T1 | TSE | 2mm | 0.3mm | None | 12cm | Cover Area of Interest |
AXIAL | T2 FS Mid TE | TSE | 2mm | 0.3mm | SPAIR | 12cm | Cover Area of Interest |
COR | T1 | TSE | 2mm | 0.3mm | None | 12cm | Angle Relative to Radio-Ulnar Joint |
COR | T2 FS Mid TE (Replace with STIR for infection) | TSE | 2mm | 0.3mm | SPAIR or STIR | 12cm | Angle Relative to Radio-Ulnar Joint |
SAG | T1 | TSE | 2mm | 0.3mm | None | 12cm | Angle Relative to Radio-Ulnar Joint |
SAG | T2 FS Mid TE (Replace with STIR for infection) | TSE | 2mm | 0.3mm | SPAIR or STIR | 12cm | Angle Relative to Radio-Ulnar Joint |
CONTRAST INJECTION | |||||||
AXIAL | T1 Fat Sat | TSE | 2mm | 0.3mm | SPIR | 12cm | Cover Area of Interest |
COR | T1 Fat Sat | TSE | 2mm | 0.3mm | SPIR | 12cm | Angle Relative to Radio-Ulnar Joint |
SAG | T1 Fat Sat | TSE | 2mm | 0.3mm | SPIR | 12cm | Angle Relative to Radio-Ulnar Joint |