MR Brain MRA WO or WWO Neuro Protocol
Scan Notes
Last updated: 3/19/2021
Charge as: MRA Brain WO or WWO
Scanner preference: 1.5T or 3T
Coil: Head
Plane | Weighting | Mode | Slice | Gap | FOV | Notes |
---|---|---|---|---|---|---|
COW | 3D TOF | 3D FFE | 1mm | 0mm | 20cm | MIP COW, Right, Left and Posterior |
Power Inject | Bolustrack. Trigger when the contrast is enters the vertebral arteries | |||||
+COW | 3D TOF | 3D FFE | 1mm | 0mm | 20cm | MIP COW, Right, Left and Posterior |
Special Technologist Instructions:
Right Click on the thumbnail and choose “VolumeView”
In the options on the top of the screen, Choose “Cut Outside”
Click on the (Star Man) icon to Generate Series
Choose the Pencil to begin Mipping your image
Orientation = "Coronal"
Stack Type = “Radial”
Radial Axis = “FH” (foot to head)
Number Of Projections = 18
Radial Angle = 10
Series Name = “COW”/”Right”/”Left”/”Basilar” whichever you are working on
Click on “Generate” to create your MIP