MR Brain WO Neuro Protocol
Problems with patient motion?
3D imaging and compressed sense are very susceptible to patient motion. If you have exhausted the usual motion-combatting strategies (padding, medication, etc.), then switch to sequences less sensitive to motion including radial k-space sampling (PROPELLER, BLADE, MultiVane, StarVibe), single-shot, or just routine 2D sequences.
Last updated: 1/21/2022
Charge as: Brain WO
Scanner preference: Not preferred on Open scanner
Coil: Head
Protocol for all scanners except MR1
Plane | Weighting | Mode | Slice (mm) | Gap (mm) | FAT SAT | FOV (cm) | MPR | Notes |
---|---|---|---|---|---|---|---|---|
AXIAL | DWI | EPI | 3 | 0.3 | YES | 23 | no | Angle to Corpus. Cover Skull Base to Vertex. Send only B1000 & ADC. |
AXIAL | T1 | 3D TFE | 1 | 0 | no | 23 | no | NO ANGLE. Cover at least 1cm above vertex through skull base. Cover entire nose. |
AXIAL | T2 | 3D TSE | 1 | 0 | no | 23 | no | NO ANGLE. Cover at least 1cm above vertex through skull base. Cover entire nose. |
AXIAL | SWI | 3D GRE | 2 | -1 | no | 23 | no | Angle to Corpus. Cover skull base to vertex. Ok to add slices. Run Ven Bold on MR1 and MR2. |
SAG | FLAIR | 3D IR-TSE | 1 | 0 | YES | 23 | no | NO ANGLE. Cover ears and nose. Spacing and gap are variable. |
Protocol for MR1
Plane | Weighting | Mode | Slice (mm) | Gap (mm) | FAT SAT | FOV (cm) | MPR | Notes |
---|---|---|---|---|---|---|---|---|
AXIAL | DWI | EPI | 3 | 0.3 | YES | 23 | no | Angle to Corpus. Cover Skull Base to Vertex. Send only B1000 & ADC. |
SAG | T1 | 3D TFE | 1 | 0 | no | 23 | no | NO ANGLE. Cover at least 1cm above vertex through skull base. Cover entire nose. |
AXIAL | T2 | 3D TSE | 2 | 0 | no | 23 | no | NO ANGLE. Cover at least 1cm above vertex through skull base. Cover entire nose. |
AXIAL | VEN BOLD | 3D GRE | 2 | -1 | no | 23 | no | Angle to Corpus. Cover skull base to vertex. Ok to add slices. Run Ven Bold on MR1 and MR2. |
SAG | FLAIR | 3D IR-TSE | 2 | -1 | YES | 23 | no | NO ANGLE. Cover ears and nose. Spacing and gap are variable. |