MR Hardware Total Spine WO or WWO

When to use OMAR/MARS:

Only use OMAR/MARS on the parts of the spine that contain hardware. If no hardware is present in a level of the spine, then use routine 2D TSE sequences instead. 

When to use fat suppression on post-gad sagittal T1:

  • Use SPIR fat suppression on post-gad sagittal T1 only if no hardware is present in that level of the spine.
  • Do not use fat suppression on post-gad sagittal T1 if hardware is present in the level of the spine you are imaging.

Other notes:

Charge as: Total Spine WO or WWO
Scanner preference: 1.5T only.
Coil: On all of our systems except MR1: Add the torso coil anteriorly in combination with the table top and/or NV array coils.

Plane Weighting Mode Slice (mm) Gap (mm) FAT SAT FOV (cm) MPR (mm) Notes
Cervical: pre-contrast
SAG T2 2D TSE 3 0.3 no no OMAR or MARS
AXIAL T2 2D TSE 3 0.3 no 14 no OMAR or MARS
SAG STIR 2D TSE 3 0.3 STIR no
SAG T1 2D TSE 3 0.3 no no
Thorcic: pre-contrast
SAG T2 2D TSE 3 0.3 no no OMAR or MARS
AXIAL T2 2D TSE 4 1 no 18 no OMAR or MARS
SAG STIR 2D TSE 3 0.3 STIR no
SAG T1 2D TSE 3 0.3 no no
Lumbar: pre-contrast
SAG T2 2D TSE 3 0.3 no no OMAR or MARS
AXIAL T2 2D TSE 4 1 no 20 no OMAR or MARS
SAG STIR 2D TSE 3 0.3 STIR no
SAG T1 2D TSE 3 0.3 no no
HAND INJECT CONTRAST
Cervical: post-contrast
SAG T1 2D TSE 3 0.3 see note no OMAR or MARS
AXIAL T1 2D TSE 3 0.3 no 14 no OMAR or MARS
Thoracic: post-contrast
SAG T1 2D TSE 3 0.3 see note no OMAR or MARS
AXIAL T1 2D TSE 4 1 no 18 no OMAR or MARS
Lumbar: post-contrast
SAG T1 2D TSE 3 0.3 see note no OMAR or MARS
AXIAL T1 2D TSE 4 1 no 20 no OMAR or MARS