MR Knee Quad Rupture WO MSK Protocol
Increase FOV to cover Quadricep Myotendoness jnction through patekkar tendon Insertion
Last updated: 4/8/19
Charge as: Knee WO
Scanner preference: 3T only
Coil: Knee Coil
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
AXIAL | T2 SPAIR | TSE | 3mm | 0.5mm | SPAIR | 20cm | Include Quad/Patellar Tendon |
COR | PD (TE=20msec) | TSE | 3mm | 0.5mm | None | 25cm | Angle to femoral condyles to patella |
COR | T2 SPAIR | TSE | 3mm | 0.5mm | SPAIR | 25cm | Angle to femoral condyles to patella |
SAG | PD (TE=20msec) | TSE | 3mm | 0.5mm | None | 25cm | Angle to femur (True Sag) |
SAG | T2 SPAIR | TSE | 3mm | 0.5mm | SPAIR | 25cm | Angle to femur (True Sag) |