MR Pediatric Abdomen WO or WWO Protocol
Scan Notes: Images must be checked by radiologist or radiology resident before giving contrast and before the patient gets off the table.
For larger patients, slice thickness may be increased up to 5mm.
Last updated:3/28/19
Charge as: Abdomen WO or WWO
Scanner preference: 1.5T or 3T
Coil: Torso or Cardiac
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
AXIAL | T1 | TSE MV MS | 3.5mm | 1mm | None | Fit to Patient | Above diaphragm to iliac crests |
AXIAL | T2 | TSE MV MS | 3.5mm | 1mm | SPAIR | Fit to Patient | Above diaphragm to iliac crests |
COR | STIR | TSE MV MS | 3.5mm | 1mm | STIR | Fit to Patient | Entire abdomen, skin to skin |
COR | DWIBS | IR EPI SS | 5mm | 1mm | None | Fit to Patient | Entire abdomen, skin to skin |
Axial | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Optional: On any patient with pancreatitis or pyelonephritis- or if requested by a radiologist.Above diaphragm to iliac crests |
COR | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Optional: On any patient with pancreatitis or pyelonephritis- or if requested by a radiologist.Entire abdomen, skin to skin |
COR | T1 | TSE MV MS | 3.5mm | 1mm | FLAIR | Fit to Patient | Optional:On any patient with pancreatitis or pyelonephritis- or if requested by a radiologist.Entire abdomen, skin to skin |
COR | T1 | TSE MV MS | 3.5mm | 1mm | None | Fit to Patient | Optional: Only if requested by a radiologist.Entire abdomen, skin to skin |
SAG | STIR | TSE MV MS | 3.5mm | 1mm | STIR | Fit to Patient | Optional: On any patient with neuroblastoma or paraspinal mass- or if requested by a radiologist.Entire abdomen, skin to skin |
If contrast is ordered: Contrast injection
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
AXIAL | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Above diaphragm to iliac crests |
COR | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Entire abdomen, skin to skin |
SAG | T1 | TSE MV MS | 3.5mm | 1mm | SPIR | Fit to Patient | Optional: On any patient with neuroblastoma or paraspinal mass- or if specifically requested by a radiologist.Entire abdomen, skin to skin |