MR Pediatric Pelvis Routine WWO Protocol

Notes

Last updated: 1/13/2022
Scanner preference: 1.5T or 3T
Coil: Torso or Cardiac

Notes from the Radiologist:

  • Power injections are preferred whenever possible.
  • Images should be checked pre-gad for sedated patients or anytime the technologist has a question/concern. For unsedated exams, no check is required.
  • For larger patients, slice thickness may be increased up to 5mm.
Plane Weighting Mode Slice Gap FS FOV Notes
AXIAL T2 TSE MV MS 3.5mm 1mm SPAIR Fit to Patient Iliac crests through pubic symphysis
COR STIR TSE MV MS 3.5mm 1mm STIR Fit to Patient skin to skin
COR DWIBS IR EPI SS 5mm 1mm None Fit to Patient skin to skin
AXIAL T1 mDIXON 3.5mm 1mm DIXON Fit to Patient skin to skin, turn ON in/out-of-phase
SAG STIR TSE MV MS 3.5mm 1mm STIR Fit to Patient Skin to Skin (make routine, previously was optional)
Inject Contrast Power injection preferred
AXIAL T1 mDIXON 3.5mm 1mm DIXON Fit to Patient Dynamic: 30, 60, 90 seconds, turn OFF in/out-of-phase
COR T1 TSE MV MS 3.5mm 1mm SPIR Fit to Patient Skin to skin
SAG T1 TSE MV MS 3.5mm 1mm SPIR Fit to Patient Skin to Skin (make routine, previously was optional)