MR Pediatric Vascular Malformation WWO Protocol

Notes

Last updated: 1/13/2022
Ordered as: MR Pediatric Vascular Malformation WWO
Scanner preference: 1.5T or 3T
Coil: Torso

Notes from the Radiologist:

  • Scan range will depend on location and extent of vascular malformation. If not specified in order, please consult radiologist after localizing lesion.
  • Images should be checked pre-gad for sedated patients or anytime the technologist has a question/concern. For unsedated exams, no check is required.
Plane Weighting Mode Slice Gap FS FOV Notes
AXIAL T2 TSE MV MS 3.5mm 1mm SPAIR Fit to Patient Radiologist to specify anatomic coverage.
COR STIR TSE MV MS 3.5mm 1mm STIR Fit to Patient Radiologist to specify anatomic coverage.
COR T1 mDIXON 3.5mm 1mm DIXON Fit to Patient Turn ON in/out-of-phase
Inject Contrast Power injection preferred
AXIAL T1 mDIXON 3.5mm 1mm DIXON Fit to patient Dynamic acquisition: 30 sec, 60 sec, 90 sec; turn OFF in/out-of-phase
COR T1 TSE MV MS 3.5mm 1mm SPIR Fit to Patient Radiologist to specify anatomic coverage.