MR TMJ WO ENT Protocol
Scan Notes:
Always Scanned Bilaterally.
Scan Right and Left sides separately.
Label Right and Left sides.
Last updated:3/28/19
Charge as: TMJ WO
Scanner preference: 1.5T preferred. Ok on 3T
Coil: Head or NV
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
AX | T1 | 4mm | 1mm | None | 23cm | Angle to Corpus- Skull Base to Vertex | 4mm |
SAG(right & left) | PD Closed | TSE | 2mm | 0mm | None | 12cm | Perpendicular to Long Axis |
SAG(right & left) | T2 Closed | TSE | 2mm | 0mm | None | 12cm | Perpendicular to Long Axis |
Coronal(right & left) | T1 Closed | TSE | 2mm | 0mm | None | 12cm | Parallel to Long Axis |
SAG(right & left) | PD Open | TSE | 2mm | 0mm | None | 12cm | Perpendicular to Long Axis |