MR Trauma CSP WO Protocol
Last updated: 10/16/2020
Charge as: CSP WO
Scanner preference: 1.5T or 3T
Coil: 16ch NV with Anterior Element (do not use top-off or 5ch spine coil in MR1 if possible)
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
SAG | T2 DRIVE | TSE | 3mm | 0.5mm | None | 20cm | Include Facet Joints |
SAG | T1 | TSE | 3mm | 0.5mm | None | 20cm | Include Facet Joints |
SAG | STIR | TSE | 3mm | 0.5mm | STIR | 20cm | Include Facet Joints |
SAG | T2- FFE | GRE | 3mm | 0.5mm | None | 20cm | Include Facet Joints |
AXIAL | T2 FAT SAT | TSE | 3mm | 0mm | SPAIR | 12cm | Above cranio-cervical junction to Mid T1 |
AXIAL | T1 | TSE | 3mm | 0mm | None | 12cm | Above cranio-cervical junction to Mid T1 |
COR | T2 FATSAT | TSE | 3mm | 0mm | SPAIR | 9cm | Cranio-cervical junction |