MRA Renal Arteries Protocol
Scan Notes
Revised 4/28/2023
Protocols are located on Siemens systems only
Plane | Weighting | Mode | Slice (mm) | Gap (mm) | FAT SAT | FOV (cm) | Notes |
---|---|---|---|---|---|---|---|
COR | T2 | HASTE BH | 4 | 0 | FIT TO PT | ||
COR | T2 | TRUFI BH | 4 | 0 | FIT TO PT | ||
AXIAL | T2 | TRUFI BH | 6 | 0 | FIT TO PT | ||
AXIAL | T1 | VIBE DIXON | 3 | 0 | FIT TO PT | GENERATE INPHASE/OUT PHASE/WATER SAT /FAT SAT SERIES | |
COR | FLASH | 3D BH | 1.1 | 0 | FIT TO PT | AP COVERAGE FROM BEHIND AORTA THROUGH ENTIRE KIDNEYS | |
Power Inject Contrast | 3 PHASE DYNAMIC | ||||||
COR | FLASH | 3D DYNAMIC BH PHASE 1 | 1.1 | 0 | FIT TO PT | AP COVERAGE FROM BEHIND AORTA THROUGH ENTIRE KIDNEYS | |
COR | FLASH | 3D DYNAMIC BH PHASE 2 | 1.1 | 0 | FIT TO PT | AP COVERAGE FROM BEHIND AORTA THROUGH ENTIRE KIDNEYS | |
COR | FLASH | 3D DYNAMIC BH PHASE 3 | 1.1 | 0 | FIT TO PT | AP COVERAGE FROM BEHIND AORTA THROUGH ENTIRE KIDNEYS | |
AXIAL | T1 | VIBE DIXON | 3 | 0 | FIT TO PT | GENERATE INPHASE/OUTPHASE/WATER SAT/FAT SAT |