MRI Pelvis W/WO MSK Protocol
Scan Notes:
If hardware is present:
- Do Axial and Coronal STIR instead of fat-sat mid-TE
- If with Contrast, do non-fat-sat T1 post-contrast
Last updated: 4/8/19
Charge as: Pelvis WWO
Scanner preference: 1.5T or 3T
Coil: Torso Coil
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
AXIAL | T1 | TSE | 5-6 mm | 1mm | None | 30cm | Entire SI Joints To Lesser Trochanters |
AXIAL | T2 SPAIR | TSE | 5-6 mm | 1mm | SPAIR | 30cm | Entire SI Joints To Lesser Trochanters |
COR | T1 | TSE | 5-6 mm | 1mm | None | 30cm | Skin to skin |
COR | T2 SPAIR | TSE | 5-6 mm | 1mm | SPAIR | 30cm | Skin to skin |
SAG | T2 SPAIR | TSE | 5-6 mm | 1mm | SPAIR | 30cm | Hip joint thru hip joint (or more, to include entire abnormality) |
Contrast Injection
Plane | Weighting | Mode | Slice | Gap | FAT SAT | FOV | Notes |
---|---|---|---|---|---|---|---|
AXIAL | T1 SPIR | TSE | 5-6 mm | 1mm | SPIR | 30cm | Same as Pre |
COR | T1 SPIR | TSE | 5-6 mm | 1mm | SPIR | 30cm | Same as Pre |
SAG | T1 SPIR | TSE | 5-6 mm | 1mm | SPIR | 30cm | Same as Pre |